5.2 Utilizing the Toolkit Software Calibration of device from Calibration Certificate
In some circumstances it may not be possible to apply inputs in which case the calibration can
be entered manually from the calibration table or certificate for a loadcell without ever having to
connect the loadcell.
Items you can change:
Number of Calibration Enter the number of points you wish to calibrate over.
Points In its simplest form you could select two for a linear
calibration. For more complex calibrations which
include linearization select three to nine points.
Input Points 1 9 Enter the input point for which you will specify a
(mV/V shown in this required engineering output value
screenshot)
Engineering Units 1 - 9 Enter the required engineering unit output for the
specified input value
Calibrate Click this button to calculate and update the device
calibration